QFM: Radiological computer-assisted prioritization software for lesions

Product code QFM covers radiological computer-assisted prioritization software for lesions, Class II devices regulated under 21 CFR 892.2080 and reviewed by FDA's Radiology panel. FDA's 510(k) database lists 41 clearances under QFM.
Updated Oct 16, 2026 · Clearances from FDA’s 510(k) database, as of Nov 16, 2026
Product code
QFM
Regulation
21 CFR 892.2080
Class
II
Panel
Radiology
Clearances
41
Most recent
Jul 8, 2026
Median review, last 2 years
126 days

What it covers

Image processing software that helps prioritize time-sensitive imaging for review by specified clinicians, based on software analysis of the images. Like QAS, it is regulated as radiological computer aided triage and notification software: it notifies, does not mark the image, does not remove cases from the reading queue, and runs in parallel with the standard of care.

Recent clearances

510(k)DeviceApplicantDecisionType
K253751CXRDetectAINeurocareai, Inc. (Dba Savelife.Ai)Jul 8, 2026Traditional
K252954MammoSightAINeurocareai, Inc. (Dba Savelife.Ai)May 4, 2026Traditional
K252482CogNet AI-MT+Medcognetics, Inc.Dec 11, 2025Traditional
K250831Annalise EnterpriseAnnalise-AiApr 23, 2025Special
K243808Rayvolve PTX-PEAZmedMar 21, 2025Traditional
K243548BriefCase-TriageAidoc Medical , Ltd.Dec 11, 2024Special
K241439VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link TriageVuno, Inc.Nov 15, 2024Traditional
K240612CINA-VCFAvicenna.AiMay 31, 2024Traditional
K233108VinDr-MammoVinbigdata Joint Stock CompanyMay 23, 2024Traditional
K232410SmartChestMilvueMay 10, 2024Traditional
From FDA’s 510(k) database, as of Nov 16, 2026Export CSV (100 rows free · all 41 with a free account)

Clearances by year

41 QFM 510(k)s were cleared from 2011 to Oct 29, 2026. Days to decision run from the day FDA received the 510(k) to its decision, including time on hold.

YearClearancesMedian days to decision
2026 (so far)2228
20253100
20246192
20238206
20229121
20215187
20204172
2019497

Who submits

27 applicants cleared a QFM device since 2011. 29% of the clearances went to applicants in United States, then Israel (24%), Australia (10%) and France (7%). By type: Traditional 90%, Special 7% and Abbreviated 2%.

ApplicantClearances since 2011
Aidoc Medical , Ltd.5
Zebra Medical Vision, Ltd.5
Annalise-Ai2
Annalise-Ai Pty , Ltd.2
Ever Fortune.Ai, Co., Ltd.2
Medcognetics, Inc.2
Neurocareai, Inc. (Dba Savelife.Ai)2
Viz. Ai, Inc.2

What the special controls require

From the special controls in 21 CFR 892.2080(b)(1), which also set labeling requirements. Read the source.

  • A detailed description of the notification and triage algorithms, their inputs and outputs, and their limitations
  • Pre-specified performance testing protocols and datasets that show effective triage, for example improved time to review of prioritized images
  • Performance results with confidence intervals across important cohorts, such as clinically relevant confounders and image acquisition characteristics
  • Standalone performance testing protocols and results
  • Software documentation: hazard analysis, requirements and design specifications, traceability, and verification and validation

Standards

21 CFR 892.2080 names no consensus standards. Check FDA’s recognized standards database, or ask which standards recent QFM clearances declared.

QFM is one of the product codes most common on FDA’s list of AI-enabled devices. See how FDA regulates AI-enabled medical devices.

Ask about QFM

US FDA devices: 510(k), De Novo, guidance, 21 CFR

Questions about QFM

What is FDA product code QFM?

Product code QFM covers radiological computer-assisted prioritization software for lesions, Class II devices regulated under 21 CFR 892.2080 and reviewed by FDA's Radiology panel. FDA's 510(k) database lists 41 clearances under QFM.

How many 510(k) clearances are there under QFM?

FDA’s 510(k) database listed 41 clearances under QFM as of Nov 16, 2026, 41 of them decided since 2011; the most recent decision was Jul 8, 2026. The Predicate Finder searches the live database.

How long does FDA take to clear a QFM 510(k)?

For the 7 QFM clearances decided in the two years before Oct 29, 2026, the median time from FDA receiving the 510(k) to its decision was 126 days. That counts calendar days, including time the submission was on hold waiting for answers from the submitter.

Which companies have the most QFM clearances?

Since 2011: Aidoc Medical , Ltd. (5), Zebra Medical Vision, Ltd. (5), Annalise-Ai (2), Annalise-Ai Pty , Ltd. (2) and Ever Fortune.Ai, Co., Ltd. (2). 27 applicants have cleared a QFM device in that time.

What testing do QFM submissions include?

From the special controls in 21 CFR 892.2080(b)(1), which also set labeling requirements. In short: a detailed description of the notification and triage algorithms, their inputs and outputs, and their limitations; pre-specified performance testing protocols and datasets that show effective triage, for example improved time to review of prioritized images; performance results with confidence intervals across important cohorts, such as clinically relevant confounders and image acquisition characteristics; standalone performance testing protocols and results; software documentation: hazard analysis, requirements and design specifications, traceability, and verification and validation.

Sources

  1. FDA product classification: QFM · Product Classification database
  2. 21 CFR 892.2080: Radiological computer aided triage and notification software · eCFR
  3. openFDA 510(k) query for QFM · run Nov 16, 2026
  4. 510(k) Premarket Notification database · FDA, accessdata.fda.gov
  5. List of Artificial Intelligence-Enabled Medical Devices · FDA, content current as of Sep 4, 2026
Follow QFMGet an email when FDA clears, recalls or publishes guidance for radiological computer-assisted prioritization software for lesions.

Know what FDA already cleared.