QBS: Radiological computer-assisted detection and diagnosis software for fracture

Product code QBS covers radiological computer-assisted detection and diagnosis software for fracture, Class II devices regulated under 21 CFR 892.2090 and reviewed by FDA's Radiology panel. FDA's 510(k) database lists 12 clearances under QBS.
Updated Oct 20, 2026 · Clearances from FDA’s 510(k) database, as of Nov 16, 2026
Product code
QBS
Regulation
21 CFR 892.2090
Class
II
Panel
Radiology
Clearances
12
Most recent
Oct 20, 2026
Median review, last 2 years
150 days

What it covers

Image processing software that helps detect, localize or characterize fractures on medical images such as radiographs, MR or CT, and tells the user about their presence, location or characteristics. Primary diagnostic and patient management decisions stay with the clinical user.

Recent clearances

510(k)DeviceApplicantDecisionType
K262062CINA-AneurysmAvicenna.AiOct 20, 2026Traditional
K261378Rayvolve AZchestAZmedAug 28, 2026Traditional
K260378RayvolveAZmedMay 12, 2026Traditional
K242171TechCare TraumaMilvueJan 17, 2025Traditional
K240712icobrain ariaIcometrix NVNov 7, 2024Traditional
K240845RayvolveAzmed SasJul 17, 2024Traditional
K223491Critical Care Suite with Pneumothorax Detection AI Algorithm, Critical Care Suite 2.1, Critical Care SuiteGe Medical Systems, LLCMay 25, 2023Traditional
K222176BoneViewGleamerMar 2, 2023Traditional
K220164RayvolveAzmed SasJun 2, 2022Traditional
K212365BoneViewGleamerMar 1, 2022Traditional
From FDA’s 510(k) database, as of Nov 16, 2026Export CSV (100 rows free · all 12 with a free account)

Clearances by year

11 QBS 510(k)s were cleared from 2011 to Oct 29, 2026. Days to decision run from the day FDA received the 510(k) to its decision, including time on hold.

YearClearancesMedian days to decision
2026 (so far)2110
20251177
20242175
20232204
20222174
20201234
20181108

Who submits

7 applicants cleared a QBS device since 2011. 64% of the clearances went to applicants in France, then United States (27%) and Belgium (9%). By type: Traditional 91% and Direct 9%.

ApplicantClearances since 2011
AZmed2
Azmed Sas2
Gleamer2
Imagen Technologies, Inc.2
Ge Medical Systems, LLC1
Icometrix NV1
Milvue1

What the special controls require

From the special controls in 21 CFR 892.2090(b)(1), as amended in June 2025, which also set labeling requirements. Read the source.

  • A detailed description of the image analysis algorithm, its inputs and outputs, and its limitations
  • Standalone performance testing with diagnostic accuracy measures such as ROC, sensitivity, specificity and predictive values, including localization accuracy when the device marks a location
  • A test dataset representative of the intended patient population, enriched only to cover important cohorts
  • Testing that shows improved assisted-read performance, such as a reader study, with readers who match the intended users
  • Software documentation, including hazard analysis, traceability, system-level test results and cybersecurity measures

Standards

21 CFR 892.2090 names no consensus standards. Check FDA’s recognized standards database, or ask which standards recent QBS clearances declared.

QBS is one of the product codes most common on FDA’s list of AI-enabled devices. See how FDA regulates AI-enabled medical devices.

Ask about QBS

US FDA devices: 510(k), De Novo, guidance, 21 CFR

Questions about QBS

What is FDA product code QBS?

Product code QBS covers radiological computer-assisted detection and diagnosis software for fracture, Class II devices regulated under 21 CFR 892.2090 and reviewed by FDA's Radiology panel. FDA's 510(k) database lists 12 clearances under QBS.

How many 510(k) clearances are there under QBS?

FDA’s 510(k) database listed 12 clearances under QBS as of Nov 16, 2026, 11 of them decided since 2011; the most recent decision was Oct 20, 2026. The Predicate Finder searches the live database.

How long does FDA take to clear a QBS 510(k)?

For the 4 QBS clearances decided in the two years before Oct 29, 2026, the median time from FDA receiving the 510(k) to its decision was 150 days. That counts calendar days, including time the submission was on hold waiting for answers from the submitter.

Which companies have the most QBS clearances?

Since 2011: AZmed (2), Azmed Sas (2), Gleamer (2), Imagen Technologies, Inc. (2) and Ge Medical Systems, LLC (1). 7 applicants have cleared a QBS device in that time.

What testing do QBS submissions include?

From the special controls in 21 CFR 892.2090(b)(1), as amended in June 2025, which also set labeling requirements. In short: a detailed description of the image analysis algorithm, its inputs and outputs, and its limitations; standalone performance testing with diagnostic accuracy measures such as ROC, sensitivity, specificity and predictive values, including localization accuracy when the device marks a location; a test dataset representative of the intended patient population, enriched only to cover important cohorts; testing that shows improved assisted-read performance, such as a reader study, with readers who match the intended users; software documentation, including hazard analysis, traceability, system-level test results and cybersecurity measures.

Sources

  1. FDA product classification: QBS · Product Classification database
  2. 21 CFR 892.2090: Radiological computer-assisted detection and diagnosis software · eCFR
  3. openFDA 510(k) query for QBS · run Nov 16, 2026
  4. 510(k) Premarket Notification database · FDA, accessdata.fda.gov
  5. List of Artificial Intelligence-Enabled Medical Devices · FDA, content current as of Sep 4, 2026
Follow QBSGet an email when FDA clears, recalls or publishes guidance for radiological computer-assisted detection and diagnosis software for fracture.