QAS: Radiological computer-assisted triage and notification software

Product code QAS covers radiological triage and notification software, Class II devices regulated under 21 CFR 892.2080 and reviewed by FDA's Radiology panel. FDA's 510(k) database lists 90 clearances under QAS.
Updated Oct 30, 2026 · Clearances from FDA’s 510(k) database, as of Nov 16, 2026
Product code
QAS
Regulation
21 CFR 892.2080
Class
II
Panel
Radiology
Clearances
90
Most recent
Oct 30, 2026
Median review, last 2 years
104 days

What it covers

Software that analyzes medical images, such as CT angiograms of the head or chest CT, and notifies a specialist of a suspected time-sensitive finding so the case can be prioritized. It runs in parallel to standard of care and does not provide a diagnosis.

Recent clearances

510(k)DeviceApplicantDecisionType
K263109EFAI NEUROSUITE CTA LVO ASSESSMENT SYSTEM (LVO-CTA-100)Ever Fortune.Ai, Co., Ltd.Oct 30, 2026Traditional
K262301Rapid ICH (v4.0)Ischemaview, Inc.Oct 9, 2026Traditional
K260906a2z-Abdo-TriageA2z Radiology Ai, Inc.Sep 16, 2026Traditional
K261085Rapid Vessel Occlusion (Rapid VO)iSchemaviewJul 13, 2026Traditional
K260729Vascular Assist Occlusion Triage (VAOT) (v1.0)GuideAI HealthJun 18, 2026Traditional
K261317BriefCase-TriageAidoc Medical , Ltd.May 14, 2026Special
K253163EFAI ERSUITE CT APPENDICITIS ASSESSMENT SYSTEM (APPEN-CT-100)Ever Fortune.Ai, Co., Ltd.Apr 22, 2026Traditional
K252421JLK-NCCTJLK, Inc.Mar 24, 2026Traditional
K253818Annalise EnterpriseHarrison-AI Medical Pty, Ltd.Mar 3, 2026Traditional
K253578BriefCase-Triage: CARE Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation deviceAidoc Medical , Ltd.Feb 26, 2026Traditional
From FDA’s 510(k) database, as of Nov 16, 2026Export CSV (100 rows free · all 90 with a free account)

Clearances by year

89 QAS 510(k)s were cleared from 2011 to Oct 29, 2026. Days to decision run from the day FDA received the 510(k) to its decision, including time on hold.

YearClearancesMedian days to decision
2026 (so far)11104
202515101
202412136
20231198
202215116
2021586
202012112
20195111
20183137

Who submits

32 applicants cleared a QAS device since 2011. 31% of the clearances went to applicants in Israel, then United States (28%), South Korea (7%) and Taiwan (7%). By type: Traditional 83%, Special 16% and Direct 1%.

ApplicantClearances since 2011
Aidoc Medical , Ltd.25
Ischemaview, Inc.8
Avicenna.Ai5
Ever Fortune.Ai, Co., Ltd.5
Annalise-Ai Pty , Ltd.4
Brainomix Limited4
JLK, Inc.4
Viz. Ai, Inc.4

Testing FDA usually expects

  • Standalone performance on a multi-site retrospective dataset kept separate from training data, reported as sensitivity and specificity with confidence intervals
  • Time to notification compared with standard of care
  • Subgroup analysis by age, sex, scanner manufacturer and acquisition protocol
  • Software verification and validation per IEC 62304, and cybersecurity documentation for connected software

Standards

QAS is one of the product codes most common on FDA’s list of AI-enabled devices. See how FDA regulates AI-enabled medical devices.

Ask about QAS

US FDA devices: 510(k), De Novo, guidance, 21 CFR

Questions about QAS

What is FDA product code QAS?

Product code QAS covers radiological triage and notification software, Class II devices regulated under 21 CFR 892.2080 and reviewed by FDA's Radiology panel. FDA's 510(k) database lists 90 clearances under QAS.

How many 510(k) clearances are there under QAS?

FDA’s 510(k) database listed 90 clearances under QAS as of Nov 16, 2026, 89 of them decided since 2011; the most recent decision was Oct 30, 2026. The Predicate Finder searches the live database.

How long does FDA take to clear a QAS 510(k)?

For the 28 QAS clearances decided in the two years before Oct 29, 2026, the median time from FDA receiving the 510(k) to its decision was 104 days. That counts calendar days, including time the submission was on hold waiting for answers from the submitter.

Which companies have the most QAS clearances?

Since 2011: Aidoc Medical , Ltd. (25), Ischemaview, Inc. (8), Avicenna.Ai (5), Ever Fortune.Ai, Co., Ltd. (5) and Annalise-Ai Pty , Ltd. (4). 32 applicants have cleared a QAS device in that time.

What testing do QAS submissions include?

Recent QAS 510(k) summaries typically describe: standalone performance on a multi-site retrospective dataset kept separate from training data, reported as sensitivity and specificity with confidence intervals; time to notification compared with standard of care; subgroup analysis by age, sex, scanner manufacturer and acquisition protocol; software verification and validation per IEC 62304, and cybersecurity documentation for connected software. Check the summaries of your own predicates for what they submitted.

Sources

  1. FDA product classification: QAS · Product Classification database
  2. 21 CFR 892.2080: Radiological computer aided triage and notification software · eCFR
  3. openFDA 510(k) query for QAS · run Nov 16, 2026
  4. 510(k) Premarket Notification database · FDA, accessdata.fda.gov
  5. List of Artificial Intelligence-Enabled Medical Devices · FDA, content current as of Sep 4, 2026