KWT: Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented
What it covers
A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
The identification in 21 CFR 888.3650, shoulder joint metal/polymer non-constrained cemented prosthesis. FDA gives KWT no definition of its own, so the regulation’s describes it.
FDA flags KWT devices as implants.
Recent clearances
| 510(k) | Device | Applicant | Decision | Type |
|---|---|---|---|---|
| K262173 | N22 Offset Humeral Head | Shoulder Innovations, Inc. | Sep 8, 2026 | Traditional |
| K250713 | Equinoxe® Humeral Reconstruction Prosthesis; Equinoxe® Stemless Shoulder Implants | Exactech, Inc. | Dec 1, 2025 | Traditional |
| K251686 | Tornier Humeral Reconstruction System Max (Tornier HRS Max) | Tornier, Inc. | Oct 30, 2025 | Traditional |
| K241944 | AGILON® XO Shoulder Replacement System | Implantcast GmbH | Mar 10, 2025 | Traditional |
| K241817 | InSet Total Shoulder System | Shoulder Innovations, Inc. | Feb 11, 2025 | Traditional |
| K231657 | AGILON® XO Shoulder Replacement System | Implantcast GmbH | Feb 16, 2024 | Traditional |
| K223631 | Comprehensive Segmental Revision System(SRS) | Biomet Orthopedics | Jul 12, 2023 | Traditional |
| K222482 | AGILON XO Shoulder System | Implantcast GmbH | Sep 28, 2022 | Traditional |
| K210899 | LINK Embrace Shoulder System- Anatomical Configuration | Waldemar Link GmbH & Co. KG | Jun 21, 2021 | Traditional |
| K191698 | TiN Coated Humeral Head | Fx Shoulder USA, Inc. | Jul 16, 2020 | Traditional |
Clearances by year
29 KWT 510(k)s were cleared from 2011 to Oct 29, 2026. Days to decision run from the day FDA received the 510(k) to its decision, including time on hold.
| Year | Clearances | Median days to decision |
|---|---|---|
| 2026 (so far) | 1 | 74 |
| 2025 | 4 | 242 |
| 2024 | 1 | 254 |
| 2023 | 1 | 219 |
| 2022 | 1 | 42 |
| 2021 | 1 | 87 |
| 2020 | 1 | 387 |
| 2019 | 4 | 91 |
| 2018 | 2 | 71 |
| 2017 | 1 | 133 |
| 2016 | 2 | 127 |
| 2015 | 2 | 98 |
| 2014 | 1 | 196 |
| 2013 | 1 | 244 |
| 2012 | 1 | 139 |
| 2011 | 5 | 165 |
Who submits
17 applicants cleared a KWT device since 2011. 69% of the clearances went to applicants in United States, then Germany (14%), Canada (7%) and France (7%). By type: Traditional 83% and Special 17%.
| Applicant | Clearances since 2011 |
|---|---|
| Catalyst Orthoscience, Inc. | 4 |
| Exactech, Inc. | 4 |
| Implantcast GmbH | 3 |
| Shoulder Innovations, Inc. | 3 |
| Fx Solutions | 2 |
| Zimmer, Inc. | 2 |
| Biomet Orthopedics | 1 |
| Biomet, Inc. | 1 |
How FDA classifies it
Class II. The special controls for this device are:
(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
From 21 CFR 888.3650. Special controls are the requirements a 510(k) must show it meets.
Standards
21 CFR 888.3650 names no consensus standards. Check FDA’s recognized standards database, or ask which standards recent KWT clearances declared.
Ask about KWT
Questions about KWT
What is FDA product code KWT?
Product code KWT, “Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented” in FDA's classification database, is a Class II device type under 21 CFR 888.3650 (shoulder joint metal/polymer non-constrained cemented prosthesis), reviewed by FDA's Orthopedic panel. FDA's 510(k) database lists 72 clearances under KWT, 29 of them since 2011.
How many 510(k) clearances are there under KWT?
FDA’s 510(k) database listed 72 clearances under KWT as of Oct 29, 2026, 29 of them decided since 2011; the most recent decision was Sep 8, 2026. The Predicate Finder searches the live database.
How long does FDA take to clear a KWT 510(k)?
For the 5 KWT clearances decided in the two years before Oct 29, 2026, the median time from FDA receiving the 510(k) to its decision was 232 days. That counts calendar days, including time the submission was on hold waiting for answers from the submitter.
Which companies have the most KWT clearances?
Since 2011: Catalyst Orthoscience, Inc. (4), Exactech, Inc. (4), Implantcast GmbH (3), Shoulder Innovations, Inc. (3) and Fx Solutions (2). 17 applicants have cleared a KWT device in that time.